• Title of article

    Optimization of two-dimensional gradient liquid chromatography separations

  • Author/Authors

    Markus Cesla، نويسنده , , Petr and H?jek، نويسنده , , Tom?? and Jandera، نويسنده , , Pavel، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    15
  • From page
    3443
  • To page
    3457
  • Abstract
    An almost orthogonal comprehensive two-dimensional liquid chromatography was developed for the separation of phenolic and flavone natural antioxidants by using combinations of a polyethylene glycol silica micro-column in the first dimension and a porous-shell fused-core C18 column in the second dimension, both in the reversed-phase mode. System orthogonality was improved using parallel gradients of acetonitrile in buffered mobile phase. A new approach was proposed to optimize matching segmented gradient profiles in the two dimensions. An algorithm was developed for automatic corrections of the shifts in retention in the second dimension induced by the parallel two-dimensional gradient operation technique. Using the porous-shell C18 column in the second dimension at elevated temperature (60 °C) and high pressure (480 bar) with optimized segmented profiles of the parallel gradients in the two dimensions, the overall separation time for comprehensive LC × LC was reduced to 30 min.
  • Keywords
    Comprehensive two-dimensional liquid chromatography , Phenolic compounds , Gradient elution , Porous-shell columns , Flavones
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2009
  • Journal title
    Journal of Chromatography A
  • Record number