Title of article :
Simultaneous determination of dimethylamine, trimethylamine and trimethylamine-n-oxide in aquatic products extracts by ion chromatography with non-suppressed conductivity detection
Author/Authors :
Li، نويسنده , , Feng and Liu، نويسنده , , Hong-ying and Xue، نويسنده , , Chang-hu and Xin، نويسنده , , Xue-qian and Xu، نويسنده , , Jun-Jie and Chang، نويسنده , , Yao-guang and Xue، نويسنده , , Yong and Yin، نويسنده , , Li-ang، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Abstract :
An ion chromatography method with non-suppressed conductivity detection was developed for the simultaneous determination of dimethylamine (DMA), trimethylamine (TMA) and trimethylamine-n-oxide (TMAO) in aquatic products. They were separated by means of cation-exchange chromatography using a 3.0 mmol/L methanesulfonic acid solution as eluent and an IonPac CS17 column (250 mm × 4 mm i.d.) as the separation column. Detection limits of dimethylamine, trimethylamine and trimethylamine-n-oxide were 0.06, 0.08 and 0.10 mg/L, respectively. The relative standard deviations (RSDs) of peak area were less than 3.53%. The recoveries were between 93.7% and 104.1%. Unlike traditional methods, this validated method is inexpensive and stable.
Keywords :
TMA , DMA , TMAO , Ion chromatography , Non-suppressed conductivity detection , Aquatic products
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A