• Title of article

    Determination of ammonium, calcium, magnesium, potassium and sodium in drinking waters by capillary zone electrophoresis on a column-coupling chip

  • Author/Authors

    Masar Gashi، نويسنده , , Mari?n and Sydes، نويسنده , , Daniel and Luc، نويسنده , , Milan and Kaniansky، نويسنده , , Du?an and Kuss، نويسنده , , Heinz-Martin، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    4
  • From page
    6252
  • To page
    6255
  • Abstract
    This work deals with simultaneous determination of ammonium, calcium, magnesium, sodium and potassium in drinking waters by capillary zone electrophoresis (CZE) on a column-coupling (CC) chip with suppressed hydrodynamic and electroosmotic transports. CZE separations were carried out in a propionate background electrolyte at a low pH (3.2) containing 18-crown-6-ether (18-crown-6) to reach a complete resolution of the cations. In addition, triethylenetetramine (TETA) coated the inner wall surface of the chip channels. The concentration limits of detection (cLOD) for the studied cations ranged from 4.9 to 11.5 μg/l concentrations using a 900 nl volume of the sample injection channel. 93–106% recoveries of the cations in drinking waters indicate a good predisposition of the present method to provide accurate analytical results.
  • Keywords
    Column-coupling chip , Capillary zone electrophoresis , Inorganic cations , Conductivity detection , Drinking water
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2009
  • Journal title
    Journal of Chromatography A
  • Record number