Title of article
Determination of ammonium, calcium, magnesium, potassium and sodium in drinking waters by capillary zone electrophoresis on a column-coupling chip
Author/Authors
Masar Gashi، نويسنده , , Mari?n and Sydes، نويسنده , , Daniel and Luc، نويسنده , , Milan and Kaniansky، نويسنده , , Du?an and Kuss، نويسنده , , Heinz-Martin، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
4
From page
6252
To page
6255
Abstract
This work deals with simultaneous determination of ammonium, calcium, magnesium, sodium and potassium in drinking waters by capillary zone electrophoresis (CZE) on a column-coupling (CC) chip with suppressed hydrodynamic and electroosmotic transports. CZE separations were carried out in a propionate background electrolyte at a low pH (3.2) containing 18-crown-6-ether (18-crown-6) to reach a complete resolution of the cations. In addition, triethylenetetramine (TETA) coated the inner wall surface of the chip channels. The concentration limits of detection (cLOD) for the studied cations ranged from 4.9 to 11.5 μg/l concentrations using a 900 nl volume of the sample injection channel. 93–106% recoveries of the cations in drinking waters indicate a good predisposition of the present method to provide accurate analytical results.
Keywords
Column-coupling chip , Capillary zone electrophoresis , Inorganic cations , Conductivity detection , Drinking water
Journal title
Journal of Chromatography A
Serial Year
2009
Journal title
Journal of Chromatography A
Record number
1512288