• Title of article

    Analysis of hydrocarbon contamination with membrane-assisted solvent extraction: Comparison of agitation and sonication methods

  • Author/Authors

    Szekeres، نويسنده , , Z. and Eke، نويسنده , , Zs. and Rikker، نويسنده , , T. and Torkos، نويسنده , , K.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    6
  • From page
    6964
  • To page
    6969
  • Abstract
    Membrane-assisted solvent extraction (MASE) coupled to large volume injection was applied to the determination of (gasoline-type) hydrocarbon contamination in water samples. Hexane was used as acceptor phase. 50 μL extract was injected in the programmed temperature vaporizer injector using combined split–splitless evaporation. The extraction conditions were optimized both for MASE with agitation and for MASE with sonication. In the course of optimization the effect of extraction time, extraction temperature, agitation speed, solvent volume, pH, ionic strength and the addition of methanol were tested. Over 75% recovery was accomplished in the range of diesel oil hydrocarbons (n-C9–n-C24). The developed method was validated. Linearity, accuracy and precision were tested. The method showed excellent linearity between 1 and 1000 μg L−1 for n-alkanes and between 0.05 and 50 mg L−1 for gasoline. The method was tested with comprehensive GC × GC as well and found to be non-discriminative to all major compounds of diesel oil.
  • Keywords
    Extractable petroleum hydrocarbon (EPH) , Hydrocarbon analysis , Membrane-assisted solvent extraction (MASE)
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2009
  • Journal title
    Journal of Chromatography A
  • Record number

    1512378