Title of article
A calculation method for the prediction of effective plate height in capillary gas chromatography
Author/Authors
Moretti، نويسنده , , P. and Vezzani، نويسنده , , S. and Castello، نويسنده , , G.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2009
Pages
6
From page
8986
To page
8991
Abstract
The effective plate height, heff, is considered to be a better measure of the efficiency of capillary column than the conventional plate height, h, in isothermal conditions. By using experimental data of 1-alcohols and n-alkanes, 2-ketones and 1-alkenes measured on capillary columns coated with non-polar stationary phases in isothermal and isobaric conditions, the peak width at half height is predicted with a function similar at that of adjusted retention time. The results obtained under different analytical conditions as the head pressure and the temperature of the column confirm the validity of the model, whose parameters are linear, and as a consequence a unique solution is obtained.
Keywords
Isothermal analysis , Gas chromatography , Effective plate height , Plate height , retention time , Peak width at half height , Capillary columns , Mathematical Models , Non-polar phases
Journal title
Journal of Chromatography A
Serial Year
2009
Journal title
Journal of Chromatography A
Record number
1512639