Title of article
Capillary isoelectric focusing coupled offline to matrix assisted laser desorption/ionization mass spectrometry
Author/Authors
Weiss، نويسنده , , Noah G. and Zwick، نويسنده , , Nicole L. and Hayes، نويسنده , , Mark A.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2010
Pages
4
From page
179
To page
182
Abstract
This work presents several critical details for making cIEF-MALDI-MS a robust technique which will allow for more routine application and aid in automation. This includes emphasis on the hardware necessary for syringe pump mobilization and proper protocol for preventing disruption from gas bubbles. Following these guidelines, excellent elution time reproducibility is demonstrated for six pI markers (RSD <5%). Additionally, the pI markers are used to calibrate the pH gradient and determine experimental pIs of proteins detected offline by mass spectrometry. This was demonstrated using a standard protein mixture of myoglobin and two forms of β-lactoglobulin. Experimental determination of protein pIs and molecular weights were found to be in agreement with literature values. The technical details discussed provide a sound foundation for applying the offline coupling of MALDI-MS with cIEF.
Keywords
MALDI-MS , MS , Syringe pump , Capillary IEF
Journal title
Journal of Chromatography A
Serial Year
2010
Journal title
Journal of Chromatography A
Record number
1512685
Link To Document