Title of article :
High-efficiency hydrophilic interaction chromatography by coupling 25 cm × 4.6 mm ID × 5 μm silica columns and operation at 80 °C
Author/Authors :
Louw، نويسنده , , Stefan and Lynen، نويسنده , , Frédéric and Hanna-Brown، نويسنده , , Melissa and Sandra، نويسنده , , Pat، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Abstract :
Recently, hydrophilic interaction chromatography (HILIC) has emerged as a valuable orthogonal tool to reversed-phase liquid chromatography (RP-LC) as it allows for resolution of highly polar ionisable compounds. The relationships between separation efficiency, column length and speed of analysis for 4.6 mm ID × 5 μm silica particle columns in HILIC are demonstrated using kinetic plots. The kinetic plots constructed for conventional pressure systems operating at 350 bar and at 30 °C and 80 °C are confirmed using experimental data for different column lengths. Efficiencies of more than 130,000 theoretical plates could be achieved by connecting up to six columns of 25 cm. As expected, a significant gain in analysis speed without loss of efficiency could be obtained by operating at 80 °C compared to 30 °C. The advantages of using long columns in HILIC in combination with elevated column temperature for the pharmaceutical industry are illustrated using test mixtures comprised of commercially available ionisable compounds (including some containing functional groups with potential genotoxic typical structural alerts) as well as real polar ionisable pharmaceuticals.
Keywords :
Hydrophilic interaction chromatography , high efficiency , Temperature , Speed , Kinetic plots
Journal title :
Journal of Chromatography A
Journal title :