• Title of article

    Determination of atracurium, cisatracurium and mivacurium with their impurities in pharmaceutical preparations by liquid chromatography with charged aerosol detection

  • Author/Authors

    B?a?ewicz، نويسنده , , Agata and Fija?ek، نويسنده , , Zbigniew and Warowna-Grze?kiewicz، نويسنده , , Ma?gorzata and Jadach، نويسنده , , Magdalena، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    7
  • From page
    1266
  • To page
    1272
  • Abstract
    The Corona CAD (charged aerosol detection) is a new type of detector introduced for LC applications that has recently become widely applied in pharmaceutical analysis. The Corona CAD measures a physical property of analyte and responds to almost all non-volatile species, independently of their nature and spectral or physicochemical properties. The LC method with charged aerosol detection was developed for the determination of three isomers of atracurium, cisatracurium and also three isomers of mivacurium with their impurities. The limit of quantitation for laudanosine was 1 μg ml−1. The elaborate method for the analysis of those active substances and laudanosine proved to be fast, precise, accurate and sensitive. All other impurities were identified using time-of-flight mass spectrometry with electrospray ionization.
  • Keywords
    Corona CAD , Charged aerosol detection , Liquid chromatography , muscle relaxants , Atracurium , Cisatracurium , Mivacurium , Laudanosine
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2010
  • Journal title
    Journal of Chromatography A
  • Record number

    1512813