Title of article
β-Diketones: Peak shape challenges and the use of mixed-mode high-performance liquid chromatography methodology to obtain fast, high resolution chromatography
Author/Authors
Rance، نويسنده , , M.J. and Wilson، نويسنده , , S.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2010
Pages
4
From page
1912
To page
1915
Abstract
Historically, indirect methods have been used for the HPLC analysis of β-diketone compounds because of the very poor peak shapes and resolution obtained on conventional HPLC stationary phases. In this paper we demonstrate that it is possible to obtain good peak shapes for underivatised β-diketone compounds, in a simulated reaction mixture, using only conventional mobile phases with mixed-mode stationary phase HPLC columns. Optimum conditions were obtained using the mixed-mode reversed-phase strong anion exchange column Primesep B, supplied by SIELC Technologies, with a 0.1% aq. TFA/MeOH gradient method and a column temperature of 55 °C.
Keywords
HPLC , Mixed-Mode , ?-diketones
Journal title
Journal of Chromatography A
Serial Year
2010
Journal title
Journal of Chromatography A
Record number
1512895
Link To Document