• Title of article

    β-Diketones: Peak shape challenges and the use of mixed-mode high-performance liquid chromatography methodology to obtain fast, high resolution chromatography

  • Author/Authors

    Rance، نويسنده , , M.J. and Wilson، نويسنده , , S.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    4
  • From page
    1912
  • To page
    1915
  • Abstract
    Historically, indirect methods have been used for the HPLC analysis of β-diketone compounds because of the very poor peak shapes and resolution obtained on conventional HPLC stationary phases. In this paper we demonstrate that it is possible to obtain good peak shapes for underivatised β-diketone compounds, in a simulated reaction mixture, using only conventional mobile phases with mixed-mode stationary phase HPLC columns. Optimum conditions were obtained using the mixed-mode reversed-phase strong anion exchange column Primesep B, supplied by SIELC Technologies, with a 0.1% aq. TFA/MeOH gradient method and a column temperature of 55 °C.
  • Keywords
    HPLC , Mixed-Mode , ?-diketones
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2010
  • Journal title
    Journal of Chromatography A
  • Record number