Title of article
Electrically assisted capillary liquid chromatography using a silica monolithic column
Author/Authors
Zhang، نويسنده , , Bo and Bergstrِm، نويسنده , , Edmund T. and Goodall، نويسنده , , David M.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2010
Pages
5
From page
2243
To page
2247
Abstract
A silica monolithic capillary column was linked to an open capillary of the same internal diameter via a Teflon sleeve to form a duplex column to investigate the combination of chromatography and electrophoresis in the mode of electrically assisted capillary liquid chromatography (eCLC). Using a commercial CE instrument with an 8.5 cm long, 100 μm i.d. reversed phase silica monolithic section and a window 1.5 cm beyond the end of this in a 21.5 cm open section, a minimum plate height of 9 μm was obtained in capillary liquid chromatography (CLC) mode at a low driving pressure of 50 psi. In eCLC mode, high speed and high resolution separations of acidic and basic compounds were achieved with selectivity tuning based on the flexible combination of pressure (0–100 psi) and voltage. Taking advantage of the excellent permeability of silica monolithic columns, use of a step flow gradient enabled elution of compounds with different charge state.
Keywords
Silica monolith , Electrically assisted , Capillary liquid chromatography , Capillary column
Journal title
Journal of Chromatography A
Serial Year
2010
Journal title
Journal of Chromatography A
Record number
1512938
Link To Document