• Title of article

    A protocol for the measurement of all the parameters of the mass transfer kinetics in columns used in liquid chromatography

  • Author/Authors

    Gritti، نويسنده , , Fabrice and Guiochon، نويسنده , , Georges، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    15
  • From page
    5137
  • To page
    5151
  • Abstract
    Band broadening in chromatography results from the combination of the dispersive effects that are associated with the different steps involved in the migration of compound bands along the column. These steps include longitudinal diffusion, trans-particle mass transfer, external film mass transfer, overall eddy diffusion, including trans-column, short-range inter-channel, trans-channel eddy diffusion, and the possible, additional mass transfer contributions arising from heat friction and the thermal heterogeneity of the column. We describe a series of experiments that provide the data needed to determine the coefficients of the contributions to band broadening of each one of these individual mass transfer steps. This specifically designed protocol can provide key information regarding the kinetic performance of columns used in liquid chromatography and explain why different columns behave so differently. The limitations, accuracy and precision of these methods are discussed. Further avenues of research that could improve the characterization of the mass transfer mechanisms in chromatographic columns, possibly contributing to the development of better columns, are suggested.
  • Keywords
    core-shell particles , Mass transfer , Liquid chromatography , Longitudinal diffusion , Eddy diffusion , External film mass transfer coefficient , Trans-particle mass transfer coefficient , Frictional heating , Totally porous particles
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2010
  • Journal title
    Journal of Chromatography A
  • Record number

    1513274