Title of article :
Hollow fiber-based liquid–liquid–liquid microextraction combined with electrospray ionization-ion mobility spectrometry for the determination of pentazocine in biological samples
Author/Authors :
Saraji، نويسنده , , Mohammad and Jafari، نويسنده , , Mohammad Taghi and Sherafatmand، نويسنده , , Hossein، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Pages :
6
From page :
5173
To page :
5178
Abstract :
A new method based on liquid–liquid–liquid microextraction combined with electrospray ionization-ion mobility spectrometry (LLLME-ESI-IMS) was used for the determination of pentazocine in urine and plasma samples. Experimental parameters which control the performance of LLLME, such as selection of composition of donor and acceptor phase, type of organic solvent, ionic strength of the sample, extraction temperature and extraction time were studied. The limit of detection and relative standard deviation of the method were 2 ng/mL and 5.3%, respectively. The linear calibration ranged from 10 to 500 ng/mL with r2 = 0.998. Pentazocine was successfully determined in urine and plasma samples without any significant matrix effect.
Keywords :
Liquid–liquid–liquid microextraction , Ion mobility spectrometry , Drug analysis , biological samples , Pentazocine