Abstract :
A method for tracking of sample components during liquid chromatography–mass spectrometry (LC–MS) method development has been proposed. The method manages to, fully automatically and without user intervention, find the chromatographic peaks in the data sets, discriminate them to sample components and track them when the separation conditions have been changed. The algorithm utilises the resolution obtained from all considered data sets and has the ability to discriminate the non informative parts. The technique has a great sensitivity even in cases where a majority of the tracked components cannot easily be spotted by means of traditional total ion chromatogram (TIC) or base peak chromatogram (BPC) representations. The method was tested on an experimental sample using six different columns and an average of 79% of the suggested sample components could be successfully tracked at a minimum area of 0.05% of the main component in the sample. 66 components with 79–92% of the total suggested component area were able to be tracked between all data sets. The method could be used to rapidly investigate selectivity during different types of separation conditions.
Keywords :
mass spectrometry , Peak detection , Chemometrics , Peak tracking , Liquid chromatography
Journal title :
Journal of Chromatography A
Serial Year :
2010
Journal title :
Journal of Chromatography A
Record number :
1513652
Link To Document :
بازگشت