• Title of article

    Direct determination of residual Pluronic F-68 in in-process samples from monoclonal antibody preparations by high performance liquid chromatography

  • Author/Authors

    Chung، نويسنده , , Hyo Helen and Zhou، نويسنده , , Chengfeng and Khor، نويسنده , , Hui Koon and Qiu، نويسنده , , Jinshu، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    8
  • From page
    2106
  • To page
    2113
  • Abstract
    A simple and sensitive high performance liquid chromatography (HPLC) method was developed to determine residual Pluronic F-68 (PF-68) in in-process samples of monoclonal antibody (MAb) preparations. The method permits the direct injection of proteinaceous samples after simple sample dilution and is able to quantitate as low as 50 mg/L of PF-68 in the presence of up to approximately 30 g/L of protein. The PF-68 molecule was separated on a restricted access reversed phase column using a step gradient and then measured by an evaporative light scattering detector (ELSD). The method was successfully applied to demonstrate PF-68 clearance in MAb purification processes. A modified colorimetric method using liquid–liquid extraction and cobalt thiocyanate to derivatize PF-68 is also described. The results obtained by both the HPLC and colorimetric methods were compared. In addition to its ease of use and simplicity, the HPLC method had better accuracy and higher throughput than the colorimetric method.
  • Keywords
    Pluronic F-68 , Restricted access media (RAM) , HPLC , Impurity clearance , Monoclonal antibody (MAb) , Colorimetric method
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2011
  • Journal title
    Journal of Chromatography A
  • Record number