Title of article
Optimization of performance of monolithic capillary column in gas chromatographic separations
Author/Authors
Korolev، نويسنده , , A. and Shyrjaeva، نويسنده , , V. and Popova، نويسنده , , T. and Kurganov، نويسنده , , A.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2011
Pages
7
From page
3267
To page
3273
Abstract
Dependence of monolithic column efficiency on column pressure was analyzed using modified Van Deemter relationship with incorporated inlet and outlet column pressures as independent variables. It was demonstrated that the highest column efficiency is observed at high pressures. Inlet and outlet pressure increase has to be controlled in such a way that the relative pressure approaches 1 and the pressure drop across the column is close to zero. Experimental results obtained for open and monolithic capillary columns confirm up to 50% higher column efficiency as compared to column efficiency under standard conditions found using conventional Van Deemter plot. Pressure increase also results in a decrease in the optimal carrier gas velocity and corresponding increase in the analysis time. This drawback can be compensated via an increase in the column temperature.
Keywords
Capillary columns , Pressure–efficiency relationship , Monolithic columns , Van Deemter relationship , Gas chromatography
Journal title
Journal of Chromatography A
Serial Year
2011
Journal title
Journal of Chromatography A
Record number
1514058
Link To Document