Title of article :
Comparison of the performance of conventional, temperature-controlled, and ultrasound-assisted ionic liquid dispersive liquid–liquid microextraction combined with high-performance liquid chromatography in analyzing pyrethroid pesticides in honey samples
Author/Authors :
Zhang، نويسنده , , Jiaheng and Gao، نويسنده , , Haixiang and Peng، نويسنده , , Bing and Li، نويسنده , , Songqing and Zhou، نويسنده , , Zhiqiang، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2011
Pages :
9
From page :
6621
To page :
6629
Abstract :
This research paper presents a comparative study of the performance of conventional, ultrasound-assisted (UA), and temperature-controlled (TC) ionic liquid (IL) dispersive liquid-phase microextraction (IL-DLLME). Various parameters that affect extraction efficiency, such as type and volume of extraction and disperser solvent, centrifugation time, salt addition, effect of temperature on TC-IL-DLLME, and effect of sonication time on UA-IL-DLLME, were evaluated. UA-IL-DLLME was found to provide the best extraction efficiency. Under optimized conditions, great enrichment factors (506–515) and good recoveries (101.2–103.0%) were obtained by analyte extraction in real samples. The limit of detections (LODs) ranged from 0.21 to 0.38 μg L−1. Good linearity was obtained in the range of 0.5–200 μg L−1 for ethofenprox and tetramethrin, and 1–200 μg L−1 for meperfluthrin and alpha-cypermethrin. Based on optimized conditions, the UA-IL-DLLME method was applied and combined with high-performance liquid chromatography with diode array detection (HPLC-DAD) to determine the presence of ethofenprox, tetramethrin, meperfluthrin, and alpha-cypermethrin in honey samples.
Keywords :
Liquid chromatography , Liquid–liquid microextraction , Ionic liquid , Ultrasound-assisted ionic liquid dispersive , Pyrethroids
Journal title :
Journal of Chromatography A
Serial Year :
2011
Journal title :
Journal of Chromatography A
Record number :
1514471
Link To Document :
بازگشت