• Title of article

    Study of the surface ionization detector for gas chromatography

  • Author/Authors

    Li، نويسنده , , Weiwei and Wu، نويسنده , , Dapeng and Chen، نويسنده , , Shiheng and Peng، نويسنده , , Hong and Guan، نويسنده , , Yafeng، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2011
  • Pages
    5
  • From page
    6812
  • To page
    6816
  • Abstract
    The structure of the surface ionization detector (SID) and the operation parameters of GC–SID were investigated to reduce peak tailing and to enhance sensitivity. The performances of the GC–SID, including its repeatability, linearity, sensitivity, selectivity, and tolerance towards water vapor, were evaluated systematically. Compared with nitrogen–phosphorus detector (NPD), the SID was able to detect fg level triethylamine, and selectively respond to alkylamines, some anilines, and some nitrogen heterocyclic compounds. Among alkylamines, the SID sensitivity to diisobutylamine was rather small. Even so, it was also still 10 times higher than that on NPD. The SID selectivity, defined as the sensitivity ratio between triethylamine and various tested non-nitrogen compounds, was higher than 106. It was found that the SID is highly tolerant towards water vapor, allowing direct injection of water sample. Finally, the GC–SID was applied to directly measure trace amines in headspace gases of rotted meat and trace simazine in tap water. The SID sensitivity to simazine was proven to be 5 times higher than that on flame ionization detector (FID). This study suggests that the SID is a promising GC detector.
  • Keywords
    Quartz liner , Amines , Selectivity , Surface ionization detector , GC , Thermo ionization
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2011
  • Journal title
    Journal of Chromatography A
  • Record number

    1514494