Title of article :
Strong cation exchange chiral stationary phase—A comparative study in high-performance liquid chromatography and subcritical fluid chromatography
Author/Authors :
Ch. Wolrab، نويسنده , , Denise and Macيkovل، نويسنده , , Pavla and Boras، نويسنده , , Mario and Kohout، نويسنده , , Michal and Lindner، نويسنده , , Wolfgang، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Pages :
8
From page :
59
To page :
66
Abstract :
The performance of a strong cation exchange-type (SCX) chiral stationary phase (CSP) was evaluated with subcritical fluid chromatography (subFC) and high performance liquid chromatography (HPLC). The chromatographic conditions in subFC were optimized by changing the amount of polar organic modifier, concentration of a basic additive in the modifier, system pressure and temperature. In this way the concentration of in situ formed transient ionic species could be varied. The gradual change of the concentration of the transient buffer, i.e. gradient elution conditions in subFC, was found beneficial for separation of a mixture of racemic compounds. The strength and amount of the in situ formed buffer was estimated on the basis of comparative experiments in subSFC and HPLC.
Keywords :
Enantiomer separation , Chiral cation exchange , Subcritical fluid chromatography , Transient buffer salts , Gradient elution , High-performance liquid chromatography
Journal title :
Journal of Chromatography A
Serial Year :
2013
Journal title :
Journal of Chromatography A
Record number :
1515307
Link To Document :
بازگشت