• Title of article

    Determination of void volume in normal phase liquid chromatography

  • Author/Authors

    Jiang، نويسنده , , Ping and Wu، نويسنده , , Di and Lucy، نويسنده , , Charles A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2014
  • Pages
    8
  • From page
    63
  • To page
    70
  • Abstract
    Void volume is an important fundamental parameter in chromatography. Little prior discussion has focused on the determination of void volume in normal phase liquid chromatography (NPLC). Various methods to estimate the total void volume are compared: pycnometry; minor disturbance method based on injection of weak solvent; tracer pulse method; hold-up volume based on unretained compounds; and accessible volume based on Martinʹs rule and its descendants. These are applied to NPLC on silica, RingSep and DNAP columns. Pycnometry provides a theoretically maximum value for the total void volume and should be performed at least once for each new column. However, pycnometry does not reflect the volume of adsorbed strong solvent on the stationary phase, and so only yields an accurate void volume for weaker mobile phase conditions. 1,3,5-Tri-t-butyl benzene (TTBB) results in hold-up volumes that are convenient measures of the void volume for all eluent conditions on charge-transfer columns (RingSep and DNAP), but is weakly retained under weak eluent conditions on silica. Injection of the weak mobile phase component (hexane) may be used to determine void volume, but care must be exercised to select the appropriate disturbance feature. Accessible volumes, that are determined using a homologous series, are always biased low, and are not recommended as a measure of the void volume.
  • Keywords
    void volume , Hold-up volume , Accessible volume , Normal phase liquid chromatography , Minor disturbance , Homologous series
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2014
  • Journal title
    Journal of Chromatography A
  • Record number

    1515590