Title of article
Rapid quantification of protein–polyethylene glycol conjugates by multivariate evaluation of chromatographic data
Author/Authors
Hansen، نويسنده , , Sigrid K. and Maiser، نويسنده , , Benjamin and Hubbuch، نويسنده , , Jürgen، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2012
Pages
7
From page
41
To page
47
Abstract
Size exclusion chromatography (SEC) is often applied for characterization of protein–polyethylene glycol (PEG) conjugates regarding the number of attached PEG chains (PEGamers). SEC analysis is advantageous as it is precise, robust, and straightforward to establish. However, most SEC based assays have a maximal throughput of a few samples per hour. We present a strategy to increase analytical throughput based on combining a short column with a fast flow rate, and finally multivariate calibration in order to compensate for the resolution lost in the trade off for speed. Different multivariate approaches were compared and multilinear regression was shown to result in the most precise calibrations. Further, a dynamic calibration approach was developed in order to account for changes in column performance over time. In this way, it was possible to establish a highly precise assay for protein PEGamer quantification with a throughput of 30 samples per hour.
Keywords
Multivariate calibration , High throughput analytics , High throughput experimentation , PEGamer quantification , Protein PEGylation , High throughput process development
Journal title
Journal of Chromatography A
Serial Year
2012
Journal title
Journal of Chromatography A
Record number
1516519
Link To Document