Title of article :
Development of an innovative and “green” stir bar sorptive extraction–thermal desorption–gas chromatography–tandem mass spectrometry method for quantification of polycyclic aromatic hydrocarbons in marine biota
Author/Authors :
Lacroix، نويسنده , , C. and Le Cuff، نويسنده , , N. and Receveur، نويسنده , , J. and Moraga، نويسنده , , D. and Auffret، نويسنده , , M. and Guyomarch، نويسنده , , J.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2014
Abstract :
There is a growing awareness of the need to reduce the negative impact of chemical analyses on the environment and to develop new eco-friendly and sustainable analytical methods without compromising performance. In this study, we developed a “green” analytical method enabling the accurate and simultaneous routine analysis of 21 polycyclic aromatic hydrocarbons (PAHs) in reduced quantities (100 mg and 1 g wet weight (WW)) of marine biota samples (fish muscle, mussel and oyster tissues) using alkaline digestion combined with stir bar sorptive extraction–thermal desorption–gas chromatography–tandem mass spectrometry (SBSE–GC–MS/MS). The innovative method provides good selectivity and specificity for most compounds. In 1 g WW samples, limits of quantification (LOQs) ranged from 1 to 10 μg/kg WW in fish muscle and from 0.5 to 10 μg/kg WW in mussel tissue. The method enables most analytes to be quantified below the restrictive limits established by the European Commission (2 and 10 μg/kg WW in fish muscle and bivalve mollusc, respectively). Higher LOQs were obtained in 100 mg WW samples ranging from 1 to 50 μg/kg WW. Recovery and linearity were assessed for all analytes. The results were satisfactory for most compounds with recoveries ranging from 94% to 117% in 1 g WW mussel samples at spike concentration of 10 ng/g WW with standard deviation not exceeding 12%. However, results confirmed that the SBSE efficiency is affected by the complexity of biological matrices, especially for high molecular weight compounds in lipid-rich mussel tissue. Because of the matrix effects, matrix-matched calibrations were carried out. Validation was performed using the standard reference material 1974c with recovery ranging from 71% to 119% except for naphthalene, anthracene and benzo(e)pyrene that were therefore not validated. Overall, the developed method meets analytical validation criteria for most compounds. Thanks to the combination of alkaline digestion and SBSE, which greatly simplifies sample treatment and limits solvent use to ethanol, the developed method followed most green analytical chemistry principles.
Keywords :
SBSE–GC–MS/MS , Marine biota , Green analytical chemistry , Pollution monitoring , PAHS
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A