Title of article :
Non-linear calibration functions in ion chromatography with suppressed conductivity detection using hydroxide eluents
Author/Authors :
Brinkmann، نويسنده , , Thomas and Specht، نويسنده , , Christian H. and Frimmel، نويسنده , , Fritz H.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2002
Abstract :
The linearity of calibration curves in ion chromatography with suppressed conductivity detection using hydroxide eluents was investigated. Theoretical calibration curves were derived for strong electrolytes and weak monobasic acids and the results compared with experimental data. At low concentrations up to 1 μmol l−1 the autoprotolysis of water induces left-curved calibration functions even for strong electrolytes like nitrate. The experimental data are best described by a quadratic function, the differences between linear and quadratic regression being up to 10%. At higher concentrations the calibration curves for strong electrolytes are linear. Due to incomplete dissociation, the calibration curves for weak mono- and dibasic acids show a right curvature. Thus, depending on the analyte and the concentration range of interest, analysts should carefully choose between a linear and a quadratic regression function.
Keywords :
nitrate , organic acids
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A