• Title of article

    Temperature-tunable selectivity in comprehensive two-dimensional gas chromatography

  • Author/Authors

    Mommers، نويسنده , , John and Knooren، نويسنده , , Jeroen and Dutriez، نويسنده , , Thomas and Ritzen، نويسنده , , Erik and van der Wal، نويسنده , , Sjoerd، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2012
  • Pages
    5
  • From page
    305
  • To page
    309
  • Abstract
    A temperature-tunable two-dimensional gas chromatography setup, consisting of three capillary columns with different selectivity, is described. In this setup the selectivity of the primary dimension can be tuned by adjusting the temperature offset of two in series-coupled capillary columns, both columns being part of the primary dimension and positioned in two separate GC ovens. The overall GC × GC separation can be optimized by altering the selectivity of the primary dimension. Besides tuning selectivity, in order to achieve optimal separation, this 2D-GC setup also offers enhanced opportunities for qualitative analysis. Sequentially altering the selectivity of the primary dimension enables one to identify groups of compounds which show similar chromatographic retention behavior.
  • Keywords
    Tunable chromatography , Tunable selectivity , ×  , GC , Dual-oven , GC 
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2012
  • Journal title
    Journal of Chromatography A
  • Record number