Title of article
Determination of carbon monoxide, methane and carbon dioxide in refinery hydrogen gases and air by gas chromatography
Author/Authors
Kami?ski، نويسنده , , Marian and Kartanowicz، نويسنده , , Rafa? and Jastrz?bski، نويسنده , , Daniel and Kami?ski، نويسنده , , Marcin M.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
7
From page
277
To page
283
Abstract
This paper illustrates a method for determining trace amounts of CO, CH4 and CO2 with the detection limit of 0.15, 0.15 and 0.20 μg/l, respectively, in refinery hydrogen gases or in air. A simple modification of a gas chromatograph equipped with a flame-ionization detector is presented. A Porapak Q column, additionally connected with a short molecular sieve 5A packed column and a catalytic hydrogenation reactor on the Ni catalyst have been applied. The principle of the analytical method proposed is the separation of CO from O2 before the introduction of CO to the methanizer. The analytical procedure and examples of the results obtained have been presented. The modification applied makes it possible to use the GC instrument for other determinations, requiring utilization of the Porapak Q column and the flame-ionization detector. In such cases, the short molecular sieve 5A column and the methanizer can be by-passed.
Keywords
CARBON MONOXIDE , Carbon dioxide , Methane , Hydrogen
Journal title
Journal of Chromatography A
Serial Year
2003
Journal title
Journal of Chromatography A
Record number
1517679
Link To Document