• Title of article

    Determination of ethylene oxide by solid-phase microextraction device with on-fiber derivatization

  • Author/Authors

    Tsai، نويسنده , , Shih-Wei and Wu، نويسنده , , Kai-Kuang Ma، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    11
  • From page
    1
  • To page
    11
  • Abstract
    The solid-phase microextraction (SPME) device was used as a time-weighted average (TWA) sampler for ethylene oxide. Carboxen/polydimethylsiloxane (CAR/PDMS) fiber was used and hydrogen bromide (HBr) was loaded onto the fiber. The SPME fiber assembly was then inserted into PTFE tubing to improve the wearer’s acceptance as a diffusive sampler. Known concentrations of ethylene oxide around the threshold limit values (TLVs)/time-weighted average and specific relative humidities (RHs) were generated by syringe pumps in a dynamic generation system. Ethylene oxide in gas bags were also generated. An exposure chamber was designed to allow measurement of face velocities, temperatures, exposing vapor concentrations, and RHs. Gas chromatography–mass spectrometry (GC–MS) was used for sample analysis. The appropriate adsorption time for SPME coating HBr was found to be 30 s and the desorption time for 2-bromothanol formed after sampling was determined to be 5 min. The experimental sampling constant of the sampler was found to be (2.96±0.09)×10−2 cm3/min, while face velocity (0–0.25 m/s) as well as RHs (10–80%) were not expected to have effects on the sampler.
  • Keywords
    ETHYLENE OXIDE
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2003
  • Journal title
    Journal of Chromatography A
  • Record number

    1517688