Title of article :
Enhanced mass resolution tandem mass spectrometry method for 2,3,7,8-tetrachlorodibenzo-p-dioxin detection with ion trap mass spectrometry using high damping gas pressure
Author/Authors :
Kemmochi، نويسنده , , Yukio and Tsutsumi، نويسنده , , Kaori and Nakazawa، نويسنده , , Hiroyuki، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
8
From page :
249
To page :
256
Abstract :
Damping gas flow was optimized for 2,3,7,8-tetrachlorodibenzo-p-dioxin (TCDD) determination using ion trap mass spectrometer. A tandem mass spectrometry (MS–MS) method with better than unit-mass resolution (mass width, 0.3 u) was developed at a damping gas flow of 1.5 ml/min and a collision-induced dissociation (CID) voltage of 3.30 V. The relative standard deviation (R.S.D.) at the enhanced resolution was 2.9% in 24 h of consecutive injections. The detection limit was significantly improved because the efficiency of both precursor ion trapping and fragmentation increased with the damping gas flow. Product ion yield was 4.5 times higher and limit of detection was 3.2 times lower than at the default flow (0.3 ml/min and 1.65 V).
Keywords :
Damping gas flow , mass spectrometry , Polychlorinated Dibenzodioxins