Title of article :
Analysis of anabolic steroids in urine by gas chromatography–microchip atmospheric pressure photoionization-mass spectrometry with chlorobenzene as dopant
Author/Authors :
Hintikka، نويسنده , , Laura and Haapala، نويسنده , , Markus and Kuuranne، نويسنده , , Tiia and Leinonen، نويسنده , , Antti and Kostiainen، نويسنده , , Risto، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Abstract :
A gas chromatography–microchip atmospheric pressure photoionization-tandem mass spectrometry (GC–μAPPI-MS/MS) method was developed for the analysis of anabolic androgenic steroids in urine as their trimethylsilyl derivatives. The method utilizes a heated nebulizer microchip in atmospheric pressure photoionization mode (μAPPI) with chlorobenzene as dopant, which provides high ionization efficiency by producing abundant radical cations with minimal fragmentation. The performance of GC–μAPPI-MS/MS was evaluated with respect to repeatability, linearity, linear range, and limit of detection (LOD). The results confirmed the potential of the method for doping control analysis of anabolic steroids. Repeatability (RSD < 10%), linearity (R2 ≥ 0.996) and sensitivity (LODs 0.05–0.1 ng/mL) were acceptable. Quantitative performance of the method was tested and compared with that of conventional GC–electron ionization-MS, and the results were in good agreement.
Keywords :
Microchip , Doping control , Anabolic Steroids , Atmospheric pressure photoionization , Tandem mass spectrometry , Gas chromatography
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A