• Title of article

    Theoretical evaluation of the advantages and limitations of constant pressure versus constant flow rate gradient elution separation in supercritical fluid chromatography

  • Author/Authors

    Pauw، نويسنده , , Ruben De and Desmet، نويسنده , , Gert and Broeckhoven، نويسنده , , Ken، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    9
  • From page
    134
  • To page
    142
  • Abstract
    This study reports on the possible advantages of switching from constant flow (cF) based gradient elution separations to constant pressure (cP) based ones in Supercritical fluid chromatography (SFC) by means of mathematical models. All the important parameters (e.g. viscosity) were fitted as a function of pressure and mol% methanol in CO2. It is shown that when switching towards a cP approach in SFC, a time gain of 34% at the end of a 0 to 50% methanol in CO2 gradient is expected. Actual analysis time gains depend on the point at which it is evaluated (e.g. up to 41% after the equilibration step). Due to the dependence of retention on density, cF- and cP-mode do not yield the same selectivity even if the same volumetric-based gradient program is used. Because the components experience a higher average pressure in the cP-mode, the apparent retention factor is slightly lower.
  • Keywords
    Kinetic performance , Supercritical fluid chromatography , Variable column length method , Constant pressure , Volume based kinetic plot method , Gradient elution
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2013
  • Journal title
    Journal of Chromatography A
  • Record number