• Title of article

    Evaluation of the pulsed discharge helium ionization detector for the analysis of hydrogen and methane in breath

  • Author/Authors

    Roberge، نويسنده , , Mark T. and Finley، نويسنده , , John W. and Lukaski، نويسنده , , Henry C. and Borgerding، نويسنده , , Anthony J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    5
  • From page
    19
  • To page
    23
  • Abstract
    Under the appropriate separation conditions the pulsed discharge helium ionization detector (PDHID) was used to detect hydrogen and methane separated from the matrix components of human breath samples. The sensitivity of this method is over an order of magnitude better than published methods using a flame ionization detector (FID) and a thermal conductivity detector (TCD), and has the further advantage of detecting both analytes with only one detector. Limits of detection were 0.3 ppmv for both hydrogen and methane and the method had a linear dynamic range (LDR) of three orders of magnitude (0.3–400 ppm, v/v). The PDHID was also compared to the FID and the TCD in regard to selectivity, sensitivity and reproducibility for high-speed gas chromatography (HSGC). It was shown that the PDHID is as sensitive as the FID for fast separations but is limited by the difficulty of resolving analyte peaks from O2 and N2. The PDHID was at least three orders of magnitude more sensitive than the TCD for all of the analytes examined.
  • Keywords
    Air analysis , Gases , Hydrogen , DetectionGC , Methane , Pulsed discharge helium ionization detection