• Title of article

    Application of a system suitability test for quality assurance and performance optimisation of a gas chromatographic system for pesticide residue analysis

  • Author/Authors

    Soboleva، نويسنده , , Eugenia and Ambrus، نويسنده , , ءrpلd، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    11
  • From page
    55
  • To page
    65
  • Abstract
    In pesticide residue analysis, screening for over 150 compounds has to be performed on a daily basis. As part of the quality control measures it is crucial to verify that the chromatographic system fits the purpose, or if any deterioration occurred during its previous use. The operation conditions of the chromatographic system can be best monitored with properly selected system suitability test (SST) mixtures, which provide information with one injection on the characteristic performance parameters of the whole system from the injector to the detectors. We developed SST mixtures that are also suitable for use with electron-capture, nitrogen–phosphorus and pulse flame photometric detectors. These SST mixtures were applied over 3 years to monitor the system performance parameters, such as the number of effective theoretical plates, resolution, asymmetry, detection limit and selectivity. The applicability and advantages of these tests are illustrated and discussed.
  • Keywords
    System suitability test , GC , Multiresidue methods , quality assurance , detection , pesticides
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2004
  • Journal title
    Journal of Chromatography A
  • Record number

    1519748