High-performance liquid chromatography–tandem mass spectrometry with a new quadrupole/linear ion trap instrument
Author/Authors
Hager، نويسنده , , James W. and Le Blanc، نويسنده , , J.C.Yves، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
7
From page
3
To page
9
Abstract
The use of a new hybrid quadrupole/linear ion trap known as the Q TRAP™ offers unique benefits as a LC–MS–MS detector for both small and large molecule analyses. The instrument combines the capabilities of a triple quadrupole mass spectrometer and ion trap technology on a single platform. Product ion scans are conducted in a hybrid fashion with the fragmentation step accomplished via acceleration into the collision cell followed by trapping and mass analysis in the Q3 linear ion trap. This results in triple quadrupole fragmentation patterns with no inherent low molecular mass cutoff. In-trap fragmentation is also possible in order to provide triple MS (MS3) capabilities. There are also several scan modes that are not possible on conventional instruments that enable identification of analytes within complex biological matrixes for subsequent high sensitivity product ion scans. This report will describe the new hybrid instrument and the principles of operation, and also provide examples of the unique scan modes and capabilities of the Q TRAP™ for LC–MS–MS detection in metabolism identification.
Keywords
mass spectrometry , instrumentation , Buspirone
Journal title
Journal of Chromatography A
Serial Year
2003
Journal title
Journal of Chromatography A
Record number
1519838
Link To Document