Title of article :
Chromatographic impulse response technique with curve fitting to measure binary diffusion coefficients and retention factors using polymer-coated capillary columns
Author/Authors :
Kong، نويسنده , , Chang Yi and Funazukuri، نويسنده , , Toshitaka and Kagei، نويسنده , , Seiichiro، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
The theoretical basis of a Gaussian-like approximate solution was applied to a chromatographic impulse response technique with curve fitting for measuring binary diffusion coefficients and retention factors using a polymer-coated capillary column. The formulae were derived for evaluating both the accuracy of the approximate solution and the sensitivity of the parameters. The validity of the solution also was confirmed experimentally for pulse injection of phenol in acetone into supercritical carbon dioxide flowing at 313.15 K and 11.6–28.6 MPa. Potential sources for experimental errors of the method are discussed.
Keywords :
Impulse response technique , Curve fitting , binary diffusion coefficient , Retention factor , transient response , Mathematical Modeling
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A