• Title of article

    Evaluation of a pulse-discharge helium ionisation detector for the determination of neon concentrations by gas chromatography

  • Author/Authors

    Lasa، نويسنده , , J and Mochalski، نويسنده , , P and Pusz، نويسنده , , J، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    4
  • From page
    261
  • To page
    264
  • Abstract
    A pulse-discharge helium ionisation detector, PDHID (Valco, PD-D2-I) with sample introduced to the discharge zone is shown to be applicable for reliable determinations of neon by gas chromatography. The detection level of 80 pg was obtained, but the dependence between detector response and neon mass was non-linear. However, for the discharge gas doped with 33 ppm of neon, a linear response to the neon mass up to 10−5 g and the detection level of 0.5 ng were obtained. The method can be used for measuring neon concentrations in groundwater systems for hydrogeological purposes.
  • Keywords
    Pulse-discharge helium ionisation detection , detection , neon , GC , Gases
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2004
  • Journal title
    Journal of Chromatography A
  • Record number