Title of article :
Influence of air on polybutadiene used in the preparation of stationary phases for high-performance liquid chromatography
Author/Authors :
Lopes، نويسنده , , Nilva P and Collins، نويسنده , , Kenneth E and Jardim، نويسنده , , Isabel C.S.F، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
6
From page :
231
To page :
236
Abstract :
A 100 ml bottle of polybutadiene, PBD, was repeatedly exposed to air over a period of 6 months. Samples were taken at time zero (PBD-0), after 3 months (PBD-3) and 6 months (PBD-6). These samples were sorbed onto HPLC silica by an open-air solution–evaporation procedure, which involved exposure to the atmosphere for 6 days. Portions of the three sets of samples were used to compare self-immobilization and the effects of 100 °C thermal treatments in air or nitrogen on HPLC performance of the resulting phases. It is concluded that self-immobilization is enhanced by prior exposure of sorbed PBD to air and subsequent heating at 100 °C further enhances column performance. The best performance (105 plates m−1) resulted from 4 h heating of PBD-6 material in nitrogen.
Keywords :
Stationary phases , LC , Polybutadiene stationary phases , Polybutadiene , thermal treatment
Journal title :
Journal of Chromatography A
Serial Year :
2004
Journal title :
Journal of Chromatography A
Record number :
1520315
Link To Document :
بازگشت