Title of article :
Coupling needle trap devices with gas chromatography–ion mobility spectrometry detection as a simple approach for on-site quantitative analysis
Author/Authors :
Nathaly Reyes-Garcés، نويسنده , , Nathaly and Gَmez-Rيos، نويسنده , , German Augusto and Souza Silva، نويسنده , , ةrica A. and Pawliszyn، نويسنده , , Janusz، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Abstract :
This study presents needle trap device (NTD) coupled with a portable gas chromatography–ion mobility spectrometry detection system (NTD-GC–IMS) as an advantageous, simple, cheap and reliable approach for on-site quantitative analysis. The NTD-GC–IMS system was evaluated using α-pinene, limonene and acetone as analytical models. Results showed satisfactory response stability (intra and inter-day relative standard deviation (RSDs) < 10%), as well as limits of detection (LODs) comparable to those provided by conventional benchtop instruments (LODs < 0.7 ng). Additionally, this methodology, together with solid phase microextraction (SPME-GC–IMS), was successfully used for on-site monitoring of temporal emissions of α-pinene by a pine branch at different times during the same day. Quantitative analysis was satisfactorily accomplished in a significantly short period of time (330 s) owing to the low detection limits that IMS detection provides.
Keywords :
volatile organic compounds , Needle trap device , Solid phase microextraction , On-site analysis , Gas chromatography–ion mobility spectrometry
Journal title :
Journal of Chromatography A
Journal title :