Author/Authors :
Mélin، نويسنده , , Carole and Lacroix، نويسنده , , Aurélie and Lalloué، نويسنده , , Fabrice and Pothier، نويسنده , , Arnaud and Zhang، نويسنده , , L.Y. and Perraud، نويسنده , , Aurélie and Dalmay، نويسنده , , Claire and Lautrette، نويسنده , , Christophe and Jauberteau، نويسنده , , Marie-Odile and Cardot، نويسنده , , Philippe and Mathonnet، نويسنده , , Muriel and Battu، نويسنده , , Serge، نويسنده ,
Abstract :
SdFFF is now commonly used for cell sorting. Nevertheless, as with many other separation methods, SdFFF Hyperlayer elution leads (1) to sample dilution resulting in cell loss which could restrict further use; and (2) to a high output flow rate impacting detector sensitivity and selectivity. In order to limit these problems, we proposed modifications of the SdFFF separation channel consisting both in downscaling and the insertion of an outlet stream splitter. This last system corresponded to a strip which divides the flow rate output into two parts, one containing concentrated cells in a reduced volume and flow rate, the other containing the excess mobile phase useless for further cell manipulation, detection and characterization. For the first time we have shown that splitter implementation and downscaling respected channel flowing and resulted in Hyperlayer elution of around 95% of cells in less than 50% of input flow rate. Improved cell sorting was demonstrated by enrichment (∼10 times) of cancer stem cells from WiDr cells with two times less quantity of injected cells.
Keywords :
Sedimentation field flow fractionation , Instrument Development , Outlet stream splitter , Channel downscaling , Cancer stem cells , Hyperlayer elution mode