Title of article :
Intermittent counter-current extraction—Equilibrium cell model, scaling and an improved bobbin design
Author/Authors :
Hewitson، نويسنده , , Peter E. Sutherland، نويسنده , , Ian and Kostanyan، نويسنده , , Artak E. and Voshkin، نويسنده , , Andrei A. and Ignatova، نويسنده , , Svetlana، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2013
Abstract :
This paper describes an equilibrium cell model for intermittent counter-current extraction that is analytically solved for the first time for continuous sample injection between a pair of columns. The model is compared with practice for injections of a model mixture of compounds on a standard high-performance counter-current chromatography instrument giving good agreement for compound elution order and the times to maximum concentration for the eluted components. An improved design of end fittings for the counter-current chromatography bobbins is described which permits on-column switching of the mobile and stationary phases. This on-column switching successfully eliminates the displaced stationary phase seen in fractions when operating ICcE with standard flying leads and gives a 6% reduction in the retention time of compounds and improved resolution due to the elimination of the time delay required to pump the previous mobile phase from standard flying leads.
Keywords :
Intermittent counter-current extraction , Flying leads , Equilibrium cell model , scale up , ICcE
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A