Title of article :
On-channel base stacking in microchip capillary gel electrophoresis for high-sensitivity DNA fragment analysis
Author/Authors :
Kim، نويسنده , , Dae Kwang and Kang، نويسنده , , Seong Ho، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
We evaluated a novel strategy for high-sensitivity DNA fragment analysis in a conventional glass double-T microfluidic chip. The microchip allows for a DNA on-channel concentration based on base stacking (BS) with a microchip capillary gel electrophoretic (MCGE) separation step in a poly(vinylpyrrolidone) (PVP) sieving matrix. Depending if low conductivity caused a neutralization reaction between the hydroxide ions and the run buffer component Tris+, the stacking of DNA fragments were processed in the microchip. Compared to a conventional MCGE separation with a normal electrokinetic injection, the peak heights of 50–2650-base pair (bp) DNA fragments on the MCGE-BS separation were increased 3.9–8.0-fold. When we applied the MCGE-BS method to the analysis of a clinical sample of bovine theileria after PCR reaction, the peak height intensity of the amplified 816-bp DNA fragment from the 18S rRNA of T. buffeli was enhanced 7.0-fold compared to that of the normal injection method.
Keywords :
Chip technology , DNA , Concentration on-channel , base stacking , stacking
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A