• Title of article

    Equilibrium in-fibre standardisation technique for solid-phase microextraction

  • Author/Authors

    Wang، نويسنده , , Yanxiang and O’Reilly، نويسنده , , John and Chen، نويسنده , , Yong and Pawliszyn، نويسنده , , Janusz، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    5
  • From page
    13
  • To page
    17
  • Abstract
    This note describes a fundamental investigation into solid-phase microextraction (SPME) using a standard loaded into the fibre coating as a means of internal standardisation for the analysis of samples contained in vials. The loading of reproducible amounts of standards into a non-porous SPME fiber was investigated. It was found that spiking low milligram quantities of standards such as benzene, toluene, ethylbenzene, xylenes (BTEX) and/or naphthalene into a few grams of pump oil sealed in a 20 mL vial provided an excellent standard generator. A single solution allowed over a hundred standard loadings with a reproducibility of <4% R.S.D. When a fiber, loaded with the standard(s) was introduced into a sample vial, extraction of analytes into the fiber and desorption of the standard(s) into the sample matrix occur simultaneously. Quantification was then based on the equilibrium distribution of the standards and the analytes between the fibre coating and the sample matrix in the vial. A comparison of equilibration profiles obtained using traditional internal standardisation and the in-fibre approach generally showed the same equilibration behaviour. The developed method was successfully used to correct for matrix effects in the BTEX analysis of a wine sample.
  • Keywords
    Internal standardisation , Solid-phase microextraction , Calibration , Automation
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2005
  • Journal title
    Journal of Chromatography A
  • Record number

    1521118