Title of article :
High speed gradient elution reversed-phase liquid chromatography
Author/Authors :
Schellinger، نويسنده , , Adam P. and Stoll، نويسنده , , Dwight R. and Carr، نويسنده , , Peter W.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Pages :
14
From page :
143
To page :
156
Abstract :
A major disadvantage of gradient elution in terms of speed results from the need to adequately re-equilibrate the column. This work distinguishes two states of re-equilibration: (1) run-to-run repeatability and (2) full equilibration. We find that excellent repeatability (±0.002 min in retention time) is achieved with at most 2 column volumes of re-equilibration whereas full equilibration can require considerably more than 20 column volumes. We have investigated the effects of adding ancillary solvents (e.g. n-propanol, n-butanol) to the eluent and changing the particle pore size, initial eluent composition and type, column temperature and flow rate on the speed of full equilibration. Full equilibration seems to be more thermodynamically limited than kinetically controlled. Also, we show that the main limitation to reducing the full equilibration time is related to instrument design issues; a novel approach to overcome these instrumental issues is described.
Keywords :
Gradient elution , equilibration , Speed , Flush-out volume
Journal title :
Journal of Chromatography A
Serial Year :
2005
Journal title :
Journal of Chromatography A
Record number :
1521251
Link To Document :
بازگشت