Title of article
Determination of iodide in seawater samples by ion chromatography with chemically-bonded poly(ethylene glycol) stationary phase
Author/Authors
Rong، نويسنده , , Li Lin Lim، نويسنده , , Lee Wah and Takeuchi، نويسنده , , Toyohide، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
5
From page
68
To page
72
Abstract
An ion chromatographic method for the rapid and direct determination of iodide in seawater is reported. Poly(ethylene glycol) (PEG) groups were chemically bonded onto silica gel or C30-bonded silica gel via diol groups. PEG-bonded C30 binary phases allowed determination of iodide in seawater samples without any interference. Effects of eluent composition on retention behavior of inorganic anions have been investigated. Both cation and anion of the eluent affected the retention of analyte anions. The retention time of anions increased with increasing eluent concentration. The detection limit for iodide obtained by injecting 0.2 μl of sample was 13 μg l−1 (S/N = 3) while the limit of quantitation was 43 μg l−1 (S/N = 10). The present method was successfully applied to the rapid and direct determination of iodide in seawater with long-term durability.
Keywords
Ion chromatography , PEG/C30 binary stationary phase , Chemically-bonded stationary phase , Seawater , iodide
Journal title
Journal of Chromatography A
Serial Year
2006
Journal title
Journal of Chromatography A
Record number
1521715
Link To Document