Title of article :
Efficacy of planar chromatography coupled to (tandem) mass spectrometry for employment in trace analysis
Author/Authors :
Jautz، نويسنده , , Ute and Morlock، نويسنده , , Gertrud، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
HPTLC/MS by a plunger-based extraction device was shown to be an appropriate technique for quantitative planar chromatography, even in trace analysis. Reproducible extraction from silica gel phases in the lower-pg range distinguishes this technique from other approaches. Repeatability of the MS signal showed a mean RSD of 12.5% for the example of Harman, a heterocyclic aromatic amine. Analytical response within a plate and over various plates/days showed determination coefficients of 0.9915 and 0.9488, respectively. Limit of detection (LOD) by a single quadrupole was better than 40 pg and limit of quantitation (LOQ) by a tandem mass spectrometer better than 20 pg. LOQ/LOD obtained were of similar magnitude as reported for HPLC/MS methods, however, this order of sensitivity was shown for the first time in the field of HPTLC/electrospray ionization MS.
Keywords :
HPTLC/ESI–MS/MS , Heterocyclic aromatic amines , Harman , Planar chromatography , HPTLC/ESI–MS , mass spectrometry
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A