Title of article :
Stereoisomer quantification of the -blocker drugs atenolol, metoprolol, and propranolol in wastewaters by chiral high-performance liquid chromatography–tandem mass spectrometry
Author/Authors :
Nikolai، نويسنده , , Lisa N. and McClure، نويسنده , , Evelyn L. and MacLeod، نويسنده , , Sherri L. and Wong، نويسنده , , Charles S.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
7
From page :
103
To page :
109
Abstract :
A chiral liquid chromatography–tandem mass spectrometry (HPLC-MS–MS) method was developed and validated for measuring individual enantiomers of three β -blocker drugs (atenolol, metoprolol, and propranolol) in wastewater treatment plant (WWTP) influents and effluents. Mean recoveries of the pharmaceuticals ranged from 67 to 106%, and the limits of detection of the analytes were 2–17 ng/L in wastewater effluents. The method was demonstrated by measuring, for the first time, the stereoisomer composition of target analytes in raw and treated wastewaters of two Canadian WWTPs. In these trials, racemic amounts of the three drugs were observed in influent of one wastewater treatment plant, but nonracemic amounts were observed in another. Effluents of the two plants contained nonracemic amounts of the drugs. These results indicate that biologically-mediated stereoselective processes that differ among WWTPs had occurred to eliminate individual enantiomers of the target analytes.
Keywords :
Chiral HPLC-MS–MS , beta-blockers , Wastewater , Enantiomers
Journal title :
Journal of Chromatography A
Serial Year :
2006
Journal title :
Journal of Chromatography A
Record number :
1521797
Link To Document :
بازگشت