Title of article :
Compact polytetrafluoroethylene assembly-type capillary electrophoresis with chemiluminescence detection
Author/Authors :
Tsukagoshi، نويسنده , , Kazuhiko and Ishida، نويسنده , , Shingo and Oda، نويسنده , , Yuichi and Noda، نويسنده , , Keiichi and Nakajima، نويسنده , , Riichiro، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
We have developed a compact polytetrafluoroethylene (PTFE) assembly-type capillary electrophoresis with chemiluminescence (CL) detection system. Luminol–microperoxidase–hydrogen peroxide chemiluminescence reaction was adopted. The device is rectangular in shape (60 mm × 40 mm × 30 mm) and includes three reservoirs (sample, migration buffer, and detection reservoirs) with electrodes. The detection reservoir includes an optical fiber to transport light at the capillary tip to a photomultiplier tube. Isoluminol isothiocyanate (ILITC) was analyzed as a model using this device with fused-silica or polytetrafluoroethylene capillary tubes 10 cm in length. We also used the sample reservoir as a reactor for an immune reaction between anti-human serum albumin immobilized on glass beads and isoluminol isothiocyanate-labeled human serum albumin. The present polytetrafluoroethylene assembly with the capillary tube was useful as a palm-sized analysis device for separation and detection, as well as a reactor.
Keywords :
Capillary electrophoresis , Chemiluminescence detection , Miniaturization , Polytetrafluoroethylene assembly
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A