Title of article
Experimental evidence of the influence of the surface chemistry of the packing material on the column pressure drop in reverse-phase liquid chromatography
Author/Authors
Gritti، نويسنده , , Fabrice and Guiochon، نويسنده , , Georges، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
10
From page
192
To page
201
Abstract
The permeabilities of six columns packed with different packing materials (neat silica, C1 endcapped silica at 3.92 μ mol/m2, C18 bonded and endcapped silica with 0.42, 1.01, 2.03, and 3.15 μ mol/m2 of C18 bonded chains) were measured. All these materials derive from the same batch of spherical particles, 5 μ m in diameter. The columns have the same tube inner diameter ( ϕ = 0.460 ± 0.003 cm) and length ( L = 15.000 ± 0.003 cm). The experimental conditions were the same, flow-rate ( F v = 1.000 ± 0.003 mL/min) and temperature (295 K). Nevertheless, it was found that the column permeability decreases significantly, by about 25%, from the neat silica column to the one packed with the highest density of C18-bonded silica (3.15 μ mol/m2). The results measured on two duplicate columns were very reproducible. Accurate ( ± 0.5 %) measurements of the hold-volumes with concentrated and dilute solutions of NO 3 − showed that the columns had all nearly the same external porosity. The result cannot be explained by the error made on the volume of the column tube either as it was measured accurately for all the columns. The residual explanation is that the interstitial velocity distribution between the packed particles depends on the chemical nature of the external surface of these particles.
Keywords
External porosity , RPLC , C18 bonding density , Total porosity , Permeability , Endcapping , Silica-C18 , Internal porosity
Journal title
Journal of Chromatography A
Serial Year
2006
Journal title
Journal of Chromatography A
Record number
1522894
Link To Document