Title of article :
Cavity ring-down spectroscopy for detection in liquid chromatography at UV wavelengths using standard cuvettes in a normal incidence geometry
Author/Authors :
L. van der Sneppen، نويسنده , , L. and Ariese، نويسنده , , F. and Gooijer، نويسنده , , C. and Ubachs، نويسنده , , W.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
Liquid chromatography (LC) with cavity ring-down spectroscopy (CRDS) detection, using flow cuvettes (put under normal incidence inside the ring-down cavity), is demonstrated. Fresnel reflections are maintained within the capture range of a stable cavity of 4 cm length. This method circumvents the need for specific Brewsterʹs angles and possible mirror degradation is avoided. The flow cuvettes are commercially available at low cost. At 355 nm (the frequency-tripled output of a Nd:YAG laser), the system surpasses the performance of conventional absorbance detectors; the baseline noise was 1.3 × 10−5 AU and detection limits (injected concentrations) were between 40 and 80 nM for nitro-polyaromatic hydrocarbons with an extinction coefficient ɛ of 7.3–10.2 × 103 M−1 cm−1. The system was also tested at 273 nm, but in the deep UV the reflectivity of the currently best available mirrors (R ≥ 99.91%) is still too low to show a significant improvement as compared to conventional UV–vis detection.
Keywords :
Cavity ring-down spectroscopy , Detection method , Flow cuvettes , Nitro-polyaromatic hydrocarbons , Liquid chromatography
Journal title :
Journal of Chromatography A
Journal title :
Journal of Chromatography A