Title of article :
Simultaneous determination of rocuronium and its eight impurities in pharmaceutical preparation using high-performance liquid chromatography with amperometric detection
Author/Authors :
B?a?ewicz، نويسنده , , Agata and Fija?ek، نويسنده , , Zbigniew and Warowna-Grze?kiewicz، نويسنده , , Ma?gorzata and Boruta، نويسنده , , Magdalena، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
7
From page :
66
To page :
72
Abstract :
A sensitive and selective HPLC method with amperometric detection (HPLC-ED) for the determination of rocuronium bromide and its eight impurities has been developed. The analysis was performed on Hypersil 100 Silica column 5 μm (250 mm × 4.6 mm; Thermo Electron). The mobile phase consisting of 4.53 g l−1 solution of tetramethylammonium hydroxide adjusted to pH 7.4 with 85% phosphoric acid:acetonitrile (1:9), was found the best for the separation and determination of the studied compounds. The chromatograms were recorded over 10 min using the amperometric detection at a potential +0.9 V of the glassy carbon electrode versus the reference electrode Ag/AgCl. The limit of quantitation was 45 ng ml−1 for rocuronium and from 25 to 750 ng ml−1 for the examined impurities. The proposed HPLC-ED method was successfully applied to the analysis of rocuronium and its impurities in Esmeron solution for injection.
Keywords :
Electrochemical detection , amperometric detection , Pharmaceutical Analysis , Rocuronium , HPLC-ED , Impurities
Journal title :
Journal of Chromatography A
Serial Year :
2007
Journal title :
Journal of Chromatography A
Record number :
1523226
Link To Document :
بازگشت