• Title of article

    Development of an accelerated low-pH reversed-phase liquid chromatography column stability test

  • Author/Authors

    Trammell، نويسنده , , Brian C. and Boissel، نويسنده , , Cheryl A. and Carignan، نويسنده , , Christina and O’Shea، نويسنده , , Daniel J. and Hudalla، نويسنده , , Christopher J. and Neue، نويسنده , , Uwe D. and Iraneta، نويسنده , , Pamela C.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    11
  • From page
    153
  • To page
    163
  • Abstract
    The important experimental design criteria for an accelerated low-pH RPLC column stability test are discussed. The influence of method variables on the amount and rate of retention-loss and the final optimized parameters for the accelerated low-pH RPLC stability test are presented. The retention-loss curves for selected C8 and C18 stationary phases are compared. These studies indicate that ligand chain length, functionality and bonding density play an important role in determining the low-pH stability of a stationary phase. Additionally, elemental analysis data are used to infer the mechanism responsible for the observed retention-loss under low-pH conditions.
  • Keywords
    LC , Stationary phases , Column stability , Bonded phases , Stability test , Reversed-phase materials
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2004
  • Journal title
    Journal of Chromatography A
  • Record number

    1523763