Title of article :
Development of an accelerated low-pH reversed-phase liquid chromatography column stability test
Author/Authors :
Trammell، نويسنده , , Brian C. and Boissel، نويسنده , , Cheryl A. and Carignan، نويسنده , , Christina and O’Shea، نويسنده , , Daniel J. and Hudalla، نويسنده , , Christopher J. and Neue، نويسنده , , Uwe D. and Iraneta، نويسنده , , Pamela C.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Pages :
11
From page :
153
To page :
163
Abstract :
The important experimental design criteria for an accelerated low-pH RPLC column stability test are discussed. The influence of method variables on the amount and rate of retention-loss and the final optimized parameters for the accelerated low-pH RPLC stability test are presented. The retention-loss curves for selected C8 and C18 stationary phases are compared. These studies indicate that ligand chain length, functionality and bonding density play an important role in determining the low-pH stability of a stationary phase. Additionally, elemental analysis data are used to infer the mechanism responsible for the observed retention-loss under low-pH conditions.
Keywords :
LC , Stationary phases , Column stability , Bonded phases , Stability test , Reversed-phase materials