    Optimization of the coupling of high-temperature liquid chromatography and flame ionization detection: Application to the separations of alcohols

  • Author/Authors

    Guillarme، نويسنده , , D. and Heinisch، نويسنده , , S. and Gauvrit، نويسنده , , J.Y. and Lanteri، نويسنده , , P. and Rocca، نويسنده , , J.L.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    6
  • From page
    22
  • To page
    27
  • Abstract
    The feasibility of coupling high-temperature liquid chromatography (HTLC) to flame ionization detection (FID) has been studied. FID parameter values (hydrogen flow-rate, air flow-rate and FID temperature), typically set in gas chromatography are rarely suitable for liquid chromatography. Best values depend obviously on the water flow rate which is defined depending on both column temperature and column internal diameter. The FID parameters were optimized according to the water flow-rate by means of an experimental design. The potential of the method is shown with some alcohol separations and the value of increasing column temperature while reducing the column diameter is highlighted.
  • Keywords
    Separation of alcohols , HTLC–FID coupling , High temperature liquid chromatography (HTLC)
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2005