Abstract :
A new HPLC method with fluorescence detection using pyridinium hydrobromide perbromide as a post-column derivatising agent has been developed to determine aflatoxin M1 in milk and cheese. The detection limits were 1 ng/kg for milk and 5 ng/kg for cheese. The calibration curve was linear from 0.001 to 0.1 ng injected. The method includes a preliminary C18-SPE clean-up and the average recoveries of Aflatoxin M1 from milk and cheese, spiked at levels of 25–75 ng/kg and 100–300 ng/kg, respectively, were 90 and 76%; the precision (RSDr) ranged from 1.7 to 2.6% for milk and from 3.5 to 6.5% for cheese. The method is rapid, easily automatable and therefore useful for accurate and precise screening of aflatoxin M1 in milk and cheese.
Keywords :
Aflatoxin M1 , milk , cheese , HPLC , Post-column derivatisation , fluorescence detection